NINIDTA, DEZCA and Maya Sari, Prima and Legiran, Legiran (2025) UJI AKURASI HASIL PENGUKURAN ITC500 PADA PEMERIKSAAN PENCITRAAN SWEPT SOURCE ANTERIOR SEGMENT OPTICAL COHERENCE TOMOGRAPHY (SS AS OCT) TERHADAP HASIL PEMERIKSAAN GONIOSKOPI PASIEN GLAUKOMA DI RSUP DR. MOHAMMAD HOESIN PALEMBANG. Masters thesis, Sriwijaya University.
Text
RAMA_11701_04032722125008.pdf - Accepted Version Restricted to Repository staff only Available under License Creative Commons Public Domain Dedication. Download (2MB) | Request a copy |
|
Text
RAMA_11701_04032722125008_TURNITIN.pdf - Accepted Version Restricted to Repository staff only Available under License Creative Commons Public Domain Dedication. Download (8MB) | Request a copy |
|
Text
RAMA_11701_04032722125008_8831030016_0018117201_01_front_ref.pdf - Accepted Version Available under License Creative Commons Public Domain Dedication. Download (490kB) |
|
Text
RAMA_11701_04032722125008_8831030016_0018117201_02.pdf - Accepted Version Restricted to Repository staff only Available under License Creative Commons Public Domain Dedication. Download (427kB) | Request a copy |
|
Text
RAMA_11701_04032722125008_8831030016_0018117201_03.pdf - Accepted Version Restricted to Repository staff only Available under License Creative Commons Public Domain Dedication. Download (148kB) | Request a copy |
|
Text
RAMA_11701_04032722125008_8831030016_0018117201_04.pdf - Accepted Version Restricted to Repository staff only Available under License Creative Commons Public Domain Dedication. Download (82kB) | Request a copy |
|
Text
RAMA_11701_04032722125008_8831030016_0018117201_05.pdf - Accepted Version Restricted to Repository staff only Available under License Creative Commons Public Domain Dedication. Download (73kB) | Request a copy |
|
Text
RAMA_11701_04032722125008_8831030016_0018117201_06.pdf - Accepted Version Restricted to Repository staff only Available under License Creative Commons Public Domain Dedication. Download (57kB) | Request a copy |
|
Text
RAMA_11701_04032722125008_8831030016_0018117201_07_ref.pdf - Bibliography Restricted to Repository staff only Available under License Creative Commons Public Domain Dedication. Download (83kB) | Request a copy |
|
Text
RAMA_11701_04032722125008_8831030016_0018117201_08_lamp.pdf - Accepted Version Restricted to Repository staff only Available under License Creative Commons Public Domain Dedication. Download (1MB) | Request a copy |
Abstract
Background: Glaucoma can be defined as progressive optic neuropathy that induces optic disc cupping and retinal ganglion cell apoptosis. Identifying angle of anterior chamber is critical to categorized glaucoma and defined its management. The current reference clinical standard for angle assessment is gonioscopy which is subjective and required patient cooperation. ITC500 as promising parameter in anterior segment optical coherence tomography has good sensitivity and specificity need further analysis in accuracy in order to be implemented in clinical practice Purpose: This study aims to test the validity, reliability, and accuracy of ITC parameter by SS AS-OCT in determining anterior chamber angle in glaucoma patients. Method: A cross-sectional study was conducted on a population of patiens with glaucoma at Glaucoma Clinic of RSUP Dr. Mohammad Hoesin starting September-December 2024. Sample collection was carried out using the consecutive sampling method and a total 120 eyes met the inclusion and exclusion criteria. Demographic characteristic data is presented in the form of a frequency distribution table. Sensitivity and specificity analysis using 2x2 cross table and bivariate analysis chi-square. Result: The majority of subject were female, with mean age of 45.53±14.37 + years old. The most common diagnosis was Primary Open Angle Glaukoma (POAG) (26.7%), diikuti dengan Ocular Hypertension (OHT) (18.3%) dan Normotension Glaucoma (NTG) (13.3%). the sensitivity result were 86.5% and specificity 98.8%, and accuracy value 95.5% to determine close angle of anterior chamber. Conclusion: ITC500 by SS AS-OCT imaging is valid, reliable, and has good accuracy for assessing anterior chamber angle.
Item Type: | Thesis (Masters) |
---|---|
Uncontrolled Keywords: | ITC500, AS-OCT, Gonioskopi, Sudut Bilik Mata Depan |
Subjects: | R Medicine > RE Ophthalmology > RE75-79 Examination. Diagnosis |
Divisions: | 04-Faculty of Medicine > 11701-Eye Diceases (Sp-1) |
Depositing User: | Dezca Nindita |
Date Deposited: | 19 Jan 2025 14:25 |
Last Modified: | 19 Jan 2025 14:25 |
URI: | http://repository.unsri.ac.id/id/eprint/165529 |
Actions (login required)
View Item |